Configure Process Measurement; Configure Sensor Flow Direction Arrow - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

4

Configure process measurement

Topics covered in this chapter:
4.1

Configure Sensor Flow Direction Arrow

Display
ProLink III
Web browser
Overview
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow
on the sensor does not match the majority of the process flow. This typically happens
when the sensor is accidentally installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output
Direction, and Totalizer Direction to control how flow is reported by the outputs and
accumulated by the totalizers and inventories.
Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display
and via digital communications. This includes ProLink III, the Web browser, and all other
user interfaces.
Configuration and Use Manual
Configure Sensor Flow Direction Arrow
Configure mass flow measurement
Configure volume flow measurement for liquid applications
Configure Gas Standard Volume (GSV) flow measurement
Configure density measurement
Configure temperature measurement
Configure Pressure Measurement Unit
Configure Velocity Measurement Unit
Menu > Configuration > Process Measurement > Flow Variables > Flow Direction
Device Tools > Configuration > Process Measurement > Flow > Sensor Direction
Configuration > Process Measurement > Flow Variables > Sensor Direction
Configure process measurement
29