Emerson Micro Motion 5700 Configuration And Use Manual page 105

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

If you want the discrete output to report a flow rate switch appropriately during fault
conditions, you must set both Discrete Output Fault Action and Process Variable Fault
Action to None.
Related information
Configuration and Use Manual
If it is set to None, and Discrete Output Source is set to Flow Rate Switch, the
discrete output will use the value determined by the current setting of Process
Variable Fault Action to determine if a flow rate switch has occurred.
If Discrete Output Source is set to any other option, the setting of Process
Variable Fault Action is irrelevant to the behavior of the discrete output during
fault conditions. The discrete output is set to the specified fault action.
Configure mA Output Fault Action
Configure Frequency Output Fault Action
Configure Discrete Output Fault Action
Configure advanced options for process measurement
97

Advertisement

Table of Contents
loading

Table of Contents