Set Up The Api Referral Application Using Prolink Iii - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configure process measurement applications
Method
Description
Digital communica-
A host writes temperature data
tions
to the meter at appropriate in-
tervals. This data will be availa-
ble in addition to the internal
temperature data.
2.
Method
Description
Digital communica-
A host writes pressure data to
tions
the meter at appropriate inter-
vals.
Postrequisites
Choose Menu > Service Tools > Service Data > View Process Variables and verify the values
for External Temperature and External Pressure.
Need help?
5.1.2

Set up the API referral application using ProLink III

This section guides you through the tasks required to set up and implement the API
referral application.
1.
2.
3.
Enable the API referral application using ProLink III
The API referral application must be enabled before you can perform any setup. If the API
referral application was enabled at the factory, you do not need to enable it now.
Prerequisites
The API referral application must be licensed on your transmitter.
58
(A, B, C, and D tables only) Choose the method to be used to supply pressure data,
and perform the required setup.
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For digital communications:
-
Verify that the host has access to the required data.
Enable the API referral application using ProLink III
Configure API referral using ProLink III
Set up temperature and pressure data for API referral using ProLink III
Setup
a. Choose Menu > Configuration > Process Measurement >
Temperature .
b. Set External Temperature to On.
c. Perform the necessary host programming and communica-
tions setup to write temperature data to the transmitter at
appropriate intervals.
Setup
a. Choose Menu > Configuration > Process Measurement >
Pressure > External Pressure .
b. Set External Pressure to On.
c. Perform the necessary host programming and communica-
tions setup to write pressure data to the transmitter at ap-
propriate intervals.
Micro Motion Model 5700 with Ethernet Transmitters

Advertisement