Measurement Performance
Figure 10-7
DeviceNet tool – Flowmeter zero procedure
10.6
Performing density calibration
Density calibration includes the following calibration points:
•
All sensors:
-
D1 calibration (low-density)
-
D2 calibration (high-density)
•
T-Series sensors only:
-
D3 calibration (optional)
-
D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibration:
•
Do not perform the D1 or D2 calibration.
•
Perform D3 calibration if you have one calibrated fluid.
•
Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and
water).
Configuration and Use Manual
Class: Calibration Object (0x65)
Instance: 1
Modify zero time if
Attribute ID: 3
required
Data type: UINT
Units: seconds
Service: Set
Class: Calibration Object (0x65)
Instance: 1
Service: 0x4B (perform or abort zero)
Perform zero
Value:
· 0: Abort zero calibration
· 1: Start zero calibration
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 16, Bit 0x8000
Check status
Value:
· 0: Zero complete
· 1: Zero in progress
Data type: USINT
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID 12, Bit 0x0020
Check outcome
Value:
· 0: Zero succeeded
· 1: Zero failed
Data type: USINT
Class: Calibration Object (0x65)
Instance: 1
Attribute ID 4: Zero standard deviation
Check zero values
Attribute ID 5: Zero offset
Data type: REAL
Units: milliseconds
103