Configure process measurement applications
Modify matrix names and labels using the a web browser
For convenience, you can change the name of a concentration matrix and the label used
for its measurement unit. This does not affect measurement.
1.
2.
3.
4.
Modify extrapolation alerts for concentration measurement
using a web browser
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These
parameters control the behavior of the concentration measurement application but do
not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature
range. If line density or line temperature goes outside the range, the transmitter will
extrapolate concentration values. However, extrapolation may affect accuracy.
Extrapolation alerts are used to notify the operator that extrapolation is occurring.
Each concentration matrix has its own extrapolation alert limits.
Procedure
1.
2.
3.
80
For digital communications:
-
Verify that the host has access to the required data.
Choose Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement page is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and
click Change Matrix.
Scroll to Step 3, then perform the following actions:
a. Set Concentration Units Label to the label that will be used for the concentration
unit.
b. If you set Concentration Units Label to Special, enter the custom label in User-
Defined Label.
c. In Matrix Name, enter the name to be used for the matrix.
Click the Apply button at the bottom of Step 3.
Choose Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and
click Change Matrix.
Scroll to Step 4.
Micro Motion Model 5700 with Ethernet Transmitters