Configure process measurement applications
2.
3.
4.
5.
Related information
Configure API referral using the display
The API referral parameters specify the API table, measurement units, and reference values
to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient
(TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1.
2.
3.
4.
56
Choose Flow Variables > Volume Flow Settings and ensure that Flow Type is set to
Liquid.
Return to the Process Measurement menu.
If the concentration measurement application is displayed in the list, choose
Concentration Measurement and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API referral application cannot
be enabled simultaneously.
Enable API referral.
a. Choose Menu > Configuration > Process Measurement > API Referral.
b. Set Enabled/Disabled to Enabled.
View the licensed features
Choose Menu > Configure > Process Measurement > API Referral.
Set API Table to the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations. Choose your API table
based on your process fluid and the measurement unit that you want to use for
referred density.
Your choice also determines the API table that will be used to calculate the
correction factor for volume (CTPL or CTL).
Refer to the API documentation and confirm your table selection.
a. Verify that your process fluid falls within range for line density, line temperature,
and line pressure.
b. Verify that the referred density range of the selected table is adequate for your
application.
If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process
fluid.
Micro Motion Model 5700 with Ethernet Transmitters