Perform A Discrete Input Loop Test Using The Display; Perform A Ma Loop Test Using Prolink Iii - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

5.
6.
7.
8.
9.
Postrequisites
15.17.4

Perform a discrete input loop test using the display

Prerequisites
Before performing a discrete input loop test, configure Channel C for a discrete input.
Follow appropriate procedures to ensure that loop testing will not interfere with existing
measurement and control loops.
Procedure
1.
2.
3.
4.
5.
Postrequisites
15.17.5

Perform a mA loop test using ProLink III

Prerequisites
Before performing a mA loop test, configure Channel C as a mA output.
Follow appropriate procedures to ensure that loop testing will not interfere with existing
measurement and control loops.
Procedure
1.
2.
3.
Configuration and Use Manual
Choose New Value.
Set Simulation Value to OFF.
Start the simulation.
Verify the signal at the receiving device.
Choose Exit.
If the discrete output readings are reversed, check the setting of Discrete Output
Polarity.
Set the remote input device to ON.
Choose Menu > Service Tools > Service Data > View I/O Levels and read the state of
the discrete input.
Set the remote input device to OFF.
Choose Menu > Service Tools > Service Data > View I/O Levels and read the state of
the discrete input.
Return the remote input device to normal operation.
If the discrete input readings are reversed, check the setting of Discrete Input
Polarity.
Choose Device Tools > Diagnostics > Testingand select the mA output to test.
Enter 4 in Fix to:.
Click Fix mA.
Troubleshooting
233