Table 15-3:
Status alerts, causes, and recommendations
Alert
Configuration and Use Manual
Name
Cause
[019] RAM Error
There is a ROM checksum
(Transmitter)
mismatch in the transmit-
ter or the RAM address lo-
cation cannot be written
in the transmitter.
This alert will not clear un-
til you reboot or power-
cycle the transmitter.
(continued)
Conditions
Recommended actions
Ensure that all wiring compartment
covers are installed correctly.
Ensure that all transmitter wiring
meets specifications and that all cable
shields are properly terminated.
Verify that the drain wires from the 4-
wire cable are properly landed.
Micro Motion recommends landing
the drain wires outside the core pro-
cessor housing. If the drain wires are
landed inside the core processor hous-
ing, they must be covered by the foil
shield for their full length, until they
land under the ground screw.
Ensure that all meter components are
grounded properly.
Evaluate the environment for sources
of high electromagnetic interference
(EMI) and relocate the transmitter or
wiring as necessary.
Reboot or power-cycle the transmitter
to see if the alert clears.
If the alert persists, replace the trans-
mitter.