Configure device options and preferences
Table 7-2:
Options for alerts and conditions
Alert
Name
[021] Incorrect Sensor Type
[030] Incorrect Board Type
Core Processor Update Failed
Password Not Set
Time Not Entered
Batcher Not Configured
[120] Curve Fit Failure (Concentration)
Bad IP network settings
110
(continued)
Conditions
Micro Motion Model 5700 with Ethernet Transmitters
Description
Transmitter verification of
sensor circuits and char-
acterization has produced
a discrepancy. The trans-
mitter cannot operate the
sensor.
The firmware or configu-
ration loaded in the trans-
mitter is incompatible
with the board type.
The core processor soft-
ware update failed.
Display security has been
enabled but the display
password has not been
changed from the default
value.
The system time has not
been entered. The system
time is required for diag-
nostic logs.
One or more of the fol-
lowing has occurred: The
batcher application is dis-
abled; No flow source has
been configured for the
batcher application; The
batch target is 0; No dis-
crete output is assigned
to batch control.
The transmitter was un-
able to calculate a valid
concentration matrix
from the current data.
IF DHCP is enabled,
the DHCP server
could not be found.
If DHCP is disabled,
there is a mismatch
between the IP ad-
dress, Subnet mask
and default gateway
settings.
Ignora-
ble
Yes
No
Yes
No
Yes
Yes
No
No