Api Referral Problems; Concentration Measurement Problems - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
15.8

API referral problems

Table 15-8:
API referral problems and recommended actions
Problem
Possible causes
Extrapolation alert is
Line pressure, line temperature, or line density
active
is outside the range of the configured API table
Inaccurate referred
density reading
15.9

Concentration measurement problems

Table 15-9:
Concentration measurement problems and recommended actions
Problem
Possible causes
Significantly incorrect
The wrong temperature or density unit was
concentration meas-
configured when the matrix was loaded
urement after loading
matrix
224
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect API table selection
Recommended actions
Check your process conditions against the
values reported by the device.
Verify the configuration of the API referral
application and related parameters.
Verify the line density value.
Verify the line temperature value.
Ensure that the application is configured to
use the appropriate temperature source.
Ensure that the pressure source is config-
ured correctly, that the external pressure
device is operating correctly, and that both
devices are using the same measurement
units.
Ensure that reference temperature and ref-
erence pressure, if applicable, are config-
ured correctly.
Ensure that the selected API table is appro-
priate for the process fluid.
Recommended actions
Set the temperature and density units to the
units used when the matrix was built, then re-
load the matrix. For custom matrices, contact
customer support.
Micro Motion Model 5700 with Ethernet Transmitters