7.2.4

Alerts, conditions, and configuration options

Table 7-2:
Options for alerts and conditions
Alert
Name
Electronics Failed
[002] RAM Error (Core Processor)
Default severi-
ty: Failure
Severity config-
[018] EEPROM Error (Transmitter)
urable: No
Fault Timeout
applicable: No
[019] RAM Error (Transmitter)
[022] Configuration Database Corrupt (Core Processor)
[024] Program Corrupt (Core Processor)
Watchdog Error
Sensor Failed
[003] Sensor Failed
Default severi-
ty: Failure
[016] Sensor Temperature (RTD) Failure
Severity config-
urable: No
Fault Timeout
[017] Sensor Case Temperature (RTD) Failure
applicable: Yes
Configuration Error
[020] Calibration Factors Missing
Default severi-
ty: Failure
Severity config-
urable: No
Fault Timeout
applicable: No
Configuration and Use Manual
Conditions
Configure device options and preferences
Description
There is an internal mem-
ory problem with the core
processor.
There is an internal mem-
ory problem with the
transmitter.
There is a ROM checksum
mismatch in the transmit-
ter or the RAM address lo-
cation cannot be written
in the transmitter.
There is a NVM checksum
mismatch in the core pro-
cessor's configuration
memory. (Standard core
processor only)
There is a checksum mis-
match in the core pro-
cessor's program section.
(Standard core processor
only)
The watchdog timer has
expired.
The pickoff amplitude is
too low.
The value computed for
the resistance of the line
RTD is outside limits
The values computed for
the resistance of the me-
ter and case RTDs are out-
side limits
Some calibration factors
have not been entered or
are incorrect.
Ignora-
ble
No
No
No
No
No
No
No
No
No
No