Configure advanced options for process measurement
4.
Example: Checking for backflow
You suspect that there is a significant amount of backflow through the sensor. To collect
data, configure two totalizers as follows:
Reset both totalizers, allow them to run for an appropriate period, then look at the amount
of reverse flow as a percentage of forward flow.
Example: Tracking three different process fluids
Three tanks are connected to a loading dock through a single meter. Each tank contains a
different process fluid. You want to track each process fluid separately.
92
Table 6-2:
Interaction between actual flow direction and Sensor Flow Direction
Arrow
Actual flow direction
Forward (same direction as
Flow arrow on sensor)
Reverse (opposite from Flow
arrow on sensor)
(Optional) Set User Name to the name you want to use for the inventory or totalizer.
User Name can have a maximum of 16 characters.
The transmitter automatically generates a name for each totalizer and inventory,
based on its source, direction, and type.
Example:
Totalizer Source=Mass Flow
Totalizer Direction=Forward Only
Totalizer name=Mass Fwd Total
Example:
Inventory Source=Gas Standard Volume Flow
Inventory Direction=Bidirectional
Inventory name=GSV Bidir Inv
The specified name is used on the transmitter display and on all interfaces that
support it. If User Name contains only spaces, the transmitter-generated name is
used. Not all interfaces support totalizer and inventory names.
Source=Mass Flow, Direction=Forward Only
Source=Mass Flow, Direction=Reverse Only
Setting of Sensor Flow Di-
rection Arrow
With Arrow
Against Arrow
With Arrow
Against Arrow
Micro Motion Model 5700 with Ethernet Transmitters
Flow direction sent to out-
puts and totalizers
Forward
Reverse
Reverse
Forward