Run an SMV test using ProLink III
1.
2.
3.
4.
5.
Postrequisites
View the results and take any appropriate actions.
Configuration and Use Manual
Read the Prerequisites in
Choose Device Tools > Diagnostics > Meter Verification > Run Test.
Tip
If Smart Meter Verification is licensed in the Model 5700 transmitter, you can also scroll to the
bottom of the ProLink III main screen and choose from the shortcut buttons: Smart Meter
Verification Overview > Meter Verification > Run Verification.
ProLink III automatically compares the contents of its SMV database to the SMV
database on the device, and uploads test data as required. You may need to wait for
a few seconds until this process is complete.
In the SMV Test Definition window, enter any desired information and click Next.
None of this information is required. It does not affect SMV processing.
ProLink III stores this information in the SMV database on the PC. It is not saved to
the transmitter.
Select the desired output behavior.
Option
Description
Continue Measuring During the test, all outputs will continue to report their assigned proc-
ess variables. The test will run for approximately 90 seconds.
Fix at Last Meas-
During the test, all outputs will report the last measured value of their
ured Value
assigned process variable. The test will run for approximately
140 seconds.
Fix at Fault
During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Click Start and wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will
return to normal behavior.
When SMV is licensed, results for the test are stored in the transmitter memory,
along with results for tests performed with any other tool. These results are also
stored in the ProLink III database. You can view and use these results in any tool-
based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but
not accessible. A Pass/Fail indication is all that is provided for the current test.