Troubleshooting
1.
2.
Important
When reassembling the meter components, be sure to grease all O-rings.
15.31
Perform a core processor resistance test
This procedure measures the resistance between the core processor terminals in the
transmitter junction box. The procedure applies only to 4-wire remote installations and
remote core processor with remote transmitter installations.
Note
Although you can perform the same test on the terminals at the core processor, the transmitter
junction box is typically easier to access.
Procedure
1.
2.
246
Plug the terminal blocks into the terminal board.
Replace the lid on the sensor junction box.
Power down the transmitter.
Remove the cover of the junction box on the transmitter to access the core
processor terminals.
Micro Motion Model 5700 with Ethernet Transmitters