6
Configure advanced options for
process measurement
Topics covered in this chapter:
6.1

Configure Response Time

Display
ProLink III
Web browser
Overview
Response Time controls the speed of various internal processes that are involved in
retrieving electronic data from the sensor and converting it to process data.
Response Time affects all process and diagnostic variables.
Restriction
Response Time is configurable only if you are using the enhanced core processor. If you are using the
standard core processor, Response Time is set to Low Filtering and cannot be changed.
Procedure
Set Response Time as desired.
Option
Normal
High Filtering Slower response. Appropriate for applications with significant amount of entrained
Low Filtering Fastest response. Appropriate for proving or filling applications.
Configuration and Use Manual
Configure Response Time
Detect and report two‐phase flow
Configure Flow Rate Switch
Configure events
Configure totalizers and inventories
Configure logging for totalizers and inventories
Configure Process Variable Fault Action
Menu > Configuration > Process Measurement > Response Time
Device Tools > Configuration > Process Measurement > Response Time
Configuration > Process Measurement > Response Time
Description
Appropriate for typical applications.