Smv History And Smv Log - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Log files, history files, and service files
Example: Historian log, 5-minute average data
S TAG:SUPPLY UID:22729F1F SW:
000000045 800:000000402
DST ON:Mountain GMT-7.0 SM:T075
SN:000000000
8/25/2014 9:58
8/25/2014 10:03
8/25/2014 10:08
8/25/2014 10:13
8/25/2014 10:18
8/25/2014 10:23
8/25/2014 10:28
8/25/2014 10:33
8/25/2014 10:38
8/25/2014 10:43
8/25/2014 10:48
8/25/2014 10:53
...
Note
The historian log is not translated. It always displays in English.
14.1.2

SMV history and SMV log

The transmitter automatically saves test data for all SMV (Smart Meter Verification) tests.
You can generate a log containing data for the 20 most recent tests or for all SMV tests.
The log is an ASCII file in .csv format.
Contents of SMV log
Each record in the SMV log represents an SMV test. Each record contains the following
information:
194
MassFlow
kg/s Max
0.0082359
0.001018
0.00099489
0.0010835
0.0011767
0.0010243
0.0010903
0.0010319
0.0011232
0.0010841
0.0010999
0.0011523
Date and time of test
Data collected during the test
The abort code (16=test completed normally)
A pass/fail result for the left pickoff (0=Pass, 1=fail)
A pass/fail result for the right pickoff (0=Pass, 1=fail)
The sensor type code
The sensor serial number
MassFlow
MassFlow
kg/s Min
kg/s Avg
0
0.00091223
0.00084441
0.00091756
0.00086279
0.00092519
0.00080879
0.00093774
0.00084206
0.00094224
0.00086888
0.00094534
0.00084823
0.00094747
0.00085327
0.00095123
0.00088614
0.00095222
0.00081306
0.00095126
0.00086106
0.00095333
0.00085537
0.00095528
Micro Motion Model 5700 with Ethernet Transmitters
MassFlow
...
kg/s Std
...
9.76E-05
...
1.61E-05
...
1.44E-05
...
2.01E-05
...
2.11E-05
...
1.85E-05
...
1.81E-05
...
1.67E-05
...
1.59E-05
...
1.99E-05
...
1.93E-05
...
2.01E-05
...