4.
5.
Example: Extrapolation alerts in action
If Extrapolation Alarm Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled,
and the active matrix is built for a temperature range of 40 °F to 80 °F, a high-temperature
extrapolation alert will be posted if line temperature goes above 82 °F.
Select the active concentration matrix using a web browser
You must select the concentration matrix to be used for measurement. Although the
transmitter can store up to six concentration matrices, only one matrix can be used for
measurement at any one time.
1.
2.
Configuration and Use Manual
Set Extrapolation Alarm Limit to the point, in percent, at which an extrapolation alert
will be posted.
Enable or disable the high and low limit alerts for temperature and density, as
desired, and click Apply.
Choose Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement page is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
Scroll to Step 2, set Active Matrix to the matrix you want to use and click Change
Matrix.
Configure process measurement applications
81