Trim Ma Output; Trim Ma Output Using The Display; Trim Ma Output Using Prolink Iii; Trim Ma Output Using A Web Browser - Emerson Micro Motion 5700 Configuration And Use Manual

Model 5700 with ethernet
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
15.18

Trim mA output

Trimming an mA output calibrates the transmitter's mA output to the receiving device. If
the current trim value is inaccurate, the transmitter will under-compensate or over-
compensate the output.
15.18.1

Trim mA output using the display

Trimming the mA output establishes a common measurement range between the
transmitter and the device that receives the mA output.
Prerequisites
Ensure that the mA output is wired to the receiving device that will be used in production.
Procedure
1.
2.
3.
15.18.2

Trim mA output using ProLink III

Trimming the mA output establishes a common measurement range between the
transmitter and the device that receives the mA output.
Prerequisites
Ensure that the mA output is wired to the receiving device that will be used in production.
Procedure
1.
2.
15.18.3

Trim mA output using a web browser

Trimming the mA output establishes a common measurement range between the
transmitter and the device that receives the mA output.
Prerequisites
Ensure that the mA output is wired to the receiving device that will be used in production.
238
Trim mA output using the display
Trim mA output using ProLink III
Trim mA output using a web browser
Choose Menu > Service Tools > mA Output Trim and select the output to trim.
Follow the instructions in the guided method.
Check the trim results. If any trim result is less than −20 microamps or greater than
+20 microamps, contact customer service.
Follow the instructions in the guided method.
Check the trim results. If any trim result is less than −20 microamps or greater than
+20 microamps, contact customer service.
(Section 15.18.1)
(Section 15.18.2)
(Section 15.18.3)