Input Configuration
Setup
Thermo Scientific
The Input Configuration Setup tab let the user assign the type of input connected to
the DensityPRO gauge and to set that input's operating range for use with additional
parameters in calculations. The user can assign analog inputs to the density,
temperature, pressure and/or flow parameters, as well as the units in which those
parameters should be displayed.
The information entered on this tab affects the information displayed on the
Additional Measurement tab. If analog input Main Board 4-20mA #1 is assigned to
Temperature Input and the Temperature Unit is set to ºF on this tab, choosing
Temperature from one of the Measurement # dropdowns on the Additional
Measurement tab will associate the input with that measurement number on the
Action and Measurement Data tabs.
While the Input Configuration tab allows the user to link four analog and
measurement inputs, there are only three measurements available on the Additional
Measurement tab.
To access the information on Application Input Configuration Setup tabs using only
the remote transmitter keypad, see
Figure 4-25. Application Screen, Input Configuration Tab
1. Determine which physical input will be connected to the device. Assign an analog
input to a measurement type by selecting the input from the Density,
Temperature, Pressure or Flow Input dropdown lists.