Filament Installation: Agilent Legacy Vs Series Helium Leak Detector; Equipment Required For Agilent Legacy Vs Series Helium Leak Detector; Installation Procedure For Agilent Legacy Vs Series Helium Leak Detector; Table 3-1. Equipment Required For Vs Models - Agilent Technologies Helium G8610 Series Field Installation Instructions

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division
3 3 3 3 Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Helium Leak Detector
Helium Leak Detector
Helium Leak Detector
Helium Leak Detector
The same filament replacement kit is compatible with any model G8610, G8611, or G8612, as well
as legacy Agilent model VS series leak detectors.
3.1 Equipment Required for Agilent Legacy VS Series
Helium Leak Detector
The following tools are suggested for assembly.
Hot Ion Source Replacement Kit (VSFLDHIS)
Extended Length M5 Allen Wrench
Metric Allen wrench set
Needle nose pliers or tweezer
M3 Phillips head screw driver
Torque wrench (must be adjustable to 40 in-
lbs (4.5 N-m) and 90 in-lbs (10.2 N-m))
Digital multimeter (Fluke 187 or equivalent)
3.2 Installation Procedure for Agilent Legacy VS Series
Helium Leak Detector
For clarity, some items have been omitted from views.
1. Turn off the power switch located on the back of the unit and unplug.
2. Wait 30 seconds for the high voltage to dissipate.
Manual 699910005, Hot Ion Source Replacement

Table 3-1. Equipment required for VS models

Item
Item
Item
Item
Figure 3-1. Rear screws
Note
Note
Note
Note
--
For removing leak detector covers
--
--
--
--
For resistance and continuity measurements
Page 16

Advertisement

Table of Contents
loading

This manual is also suitable for:

Helium g8611 seriesHelium g8612 seriesHelium vs series

Table of Contents