Vacuum Products Division
3 3 3 3 Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Filament Installation: Agilent Legacy VS Series
Helium Leak Detector
Helium Leak Detector
Helium Leak Detector
Helium Leak Detector
The same filament replacement kit is compatible with any model G8610, G8611, or G8612, as well
as legacy Agilent model VS series leak detectors.
3.1 Equipment Required for Agilent Legacy VS Series
Helium Leak Detector
The following tools are suggested for assembly.
Hot Ion Source Replacement Kit (VSFLDHIS)
Extended Length M5 Allen Wrench
Metric Allen wrench set
Needle nose pliers or tweezer
M3 Phillips head screw driver
Torque wrench (must be adjustable to 40 in-
lbs (4.5 N-m) and 90 in-lbs (10.2 N-m))
Digital multimeter (Fluke 187 or equivalent)
3.2 Installation Procedure for Agilent Legacy VS Series
Helium Leak Detector
For clarity, some items have been omitted from views.
1. Turn off the power switch located on the back of the unit and unplug.
2. Wait 30 seconds for the high voltage to dissipate.
Manual 699910005, Hot Ion Source Replacement
Table 3-1. Equipment required for VS models
Item
Item
Item
Item
Figure 3-1. Rear screws
Note
Note
Note
Note
--
For removing leak detector covers
--
--
--
--
For resistance and continuity measurements