VS Series Helium Mass Spectrometer Leak Detectors
Section 1. Introduction to the VS Series

1.1 The VS Series Leak Detector

The VS Series is a wide-range Helium Mass Spectrometer Leak Detector, which is
comprised of a turbo molecular high vacuum pump, internal mechanical rough pump
(model VS PR02), internal scroll pump (model PD03), external mechanical rough pump
(models VS MR15, VS MD15, VS MD30+, VS BR15, VS BD15 and VS BD30+),
spectrometer tube, valve block, leak detector electronics, an operator interface (TFT-LCD
display with a touch panel and two hard buttons: TEST and VENT), and optional features,
including:
Wireless remote control that includes:
Optional I/O PCB (P/N VSFLDIO).
Models VS MR15x, VS MD15x and VS MD30+x include an optional two-wheel cart.
All models are available in 100/115 VAC and 230 VAC configurations (see Table 1-1 on
page 1-2).
The wireless remote control has the following capabilities:
RF communication between VS leak detector (Base Station or Host) and the Wireless
Remote unit from the distance up to 100 m (300 ft).
Enter the VS leak detector unit into Test and Hold modes from the Wireless Remote
unit keypad.
Read the VS leak detector unit's leak rate in a linear and logarithmic bar graph
format with a numeric range indicator and System status in alphanumeric format on
the Wireless Remote unit LCD display.
Configure up to ten different VS leak detector units for wireless control by toggling
through a MAC address list using one Wireless Remote unit.
The optional I/O PCB compromises:
Optically isolated parallel I/O (DB25 female connector with 12 output and 8 input
lines) to use with an external PLC and should be powered by the user (+5-24 VDC)
via a cable with length less than 10' and provided by the user (see Appendix A.3.2
"I/O Connector" on page A-14).
Leak Detector
Wireless Base PCB (P/N G8600-68001) installed in a VS leak detector
Wireless Remote unit (P/N G8600-68002).
1-1