Voltage Source Calibration Commands; Humidity Calibration Commands - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 15: Calibration procedure

Voltage source calibration commands

This is a list of all voltage source calibration commands for the Model 6517B.
Voltage source calibration commands
Command
:CAL:PROT:VSETZ100
:CAL:PROT:VSRCZ100 <measured voltage>
:CAL:PROT:VSET100
:CAL:PROT:VSRC100 <measured voltage>
:CAL:PROT:VSETN100
:CAL:PROT:VSRCN100 <measured voltage>
:CAL:PROT:VSET40
:CAL:PROT:VSRC40 <measured voltage>
:CAL:PROT:VSETZ1000
:CAL:PROT:VSRCZ1000 <measured voltage>
:CAL:PROT:VSET1000
:CAL:PROT:VSRC1000 <measured voltage>
:CAL:PROT:VSETN1000
:CAL:PROT:VSRCN1000 <measured voltage>
:CAL:PROT:VSET400
:CAL:PROT:VSRC400 <measured voltage>

Humidity calibration commands

This is a list of all Model 6517B humidity calibration commands
Humidity calibration commands
Command
:CAL:PROT:HUMZERO
:CAL:PROT:HUM1 <1 volt value>
15-22
Model 6517B Electrometer Reference Manual
Description
Start the 100 V range zero calibration
Calibrate the 100 V range zero
Start the +100 V range calibration
Calibrate the +100 V range
Start the -100 V range calibration
Calibrate the -100 V range
Start the +40 V calibration
Calibrate the +40 V output
Start the 1000 V range zero calibration
Calibrate the 1000 V range zero
Start the +1000 V range calibration
Calibrate the +1000 V range
Start the -1000 V range calibration
Calibrate the -1000 V range
Start the +400 V calibration
Calibrate the +400 V output
Description
Calibrate the humidity input zero
Calibrate the humidity +1 V range
6517B-901-01 Rev. C / August 2015