Llockout <B; Basic Trigger Commands; [:Execute] - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference
:LLOCkout <b>
This command enables or disables local lockout.
Usage
:SYSTem:LLOCkout <b>
:LLOCkout?
<b>
Details
If enabled, the front-panel keys are locked out (non-operational) when the instrument is in remote. If
disabled, the front panel keys are operational in remote. The instrument must be in remote in order to
use this command.
Taking the instrument out of remote restores front-panel key operation, but does not change the
status of the :LLOCkout command.
Note that this command can only be sent over the RS-232 interface.
Also see
:LOCal
(on page 14-133)
:REMote
(on page 14-133)

Basic trigger commands

The following commands are used to configure the Model 6517B for basic triggering. See
(on page 7-1) for details on basic triggering.

[:EXECute]

This command exits from advanced triggering and selects basic triggering.
Usage
:SYSTem:MACRo:TRIGger:MODE
Details
When this action command is sent, the instrument exits from advanced triggering and select basic
triggering. Sending this command while already in basic triggering has no effect on the instrument.
Also see
None
14-134
0 or OFF: Disable local lockout
1 or ON: Enable local lockout
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015