Model 6517B Electrometer Reference Manual
Measurement event
Reading overflow
Low limit 1
High limit 1
Low limit 2
High limit 2
Reading available
Reading underflow
Buffer available
Buffer half-full
Buffer full
Sequence reading
available
Buffer pretriggered
Out of limits error
Fixture lid closed
V-source compliance
Negative transition effect on questionable event register
Questionable event
Volts summary
Amps summary
Temperature summary
Humidity summary
Ohms summary
Coulombs summary
Sequence test summary Sets B12 when the sequence test is re-started
Negative transition effect on operation event register
Operation event
Calibrating
Trigger layer
Arm layer
Calculating
Idle
Sequence test running
Negative transition effect on trigger event register
Trigger event
Sequence 1
Negative transition effect on arm event register
Arm event
6517B-901-01 Rev. C / August 2015
Negative transition effect on measurement event register
Sets B0 when a reading has gone from overflow to normal
Sets B1 when the reading is no longer less than the low limit 1 setting
Sets B2 when the reading is no longer greater than the low limit 1 setting
Sets B3 when the reading is no longer less than the low limit 2 setting
Sets B4 when the reading is no longer greater than the high limit 2 setting
Sets B5 when the next reading is being taken
Sets B6 when the reading underflow condition has cleared
Sets B7 when the trace buffer has become empty
Sets B8 when the trace buffer is no longer half-full
Sets B9 when the trace buffer is no longer full
Sets B10 when an alternating polarity measurement is being read
Sets B11 when the trace buffer pretrigger operations are completed
Sets B12 when the out of limits error has cleared
Sets B13 when fixture lid is opened
Sets B14 when V-source comes out of compliance
Negative transition effect on questionable event register
Sets B0 when a subsequent valid volts measurement occurs
Sets B1 when a subsequent valid amps measurement occurs
Sets B4 when a subsequent valid external temperature reading occurs
Sets B9 when a subsequent valid relative humidity reading occurs
Sets B10 when a subsequent valid ohms measurement occurs
Sets B11 when a subsequent valid coulombs measurement occurs
Negative transition effect on operation event register
Sets B0 at the end of calibration
Sets B5 when leaving the trigger layer
Sets B6 when leaving an arm layer
Sets B9 when the reading calculation is completed
Sets B10 when leaving the idle state
Sets B11 when the aborted test is restarted
Negative transition effect on trigger event register
Sets B1 when leaving the trigger layer
Negative transition effect on arm event register
Section 14: SCPI command reference
14-117