Immediate; Data; Calculate3 - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference

:IMMediate

This command is used to perform CALC 2.
Usage
:CALCulate2:IMMediate
:IMMediate?
Details
The :IMMediate command is used to perform the selected CALC 2 operation on the readings in the
buffer (assuming CALC 2 is enabled). After performing the calculation, the result can be read by
using the :CALCulate2:DATA? query command.
An alternate way to perform the calculation and read the result is by using the query form of the
command (:IMMediate?). When this command is sent, the calculation is performed and the result is
queried.
Also see
None

DATA?

This command is used to read the CALC 2 result.
Usage
:CALCulate2:DATA?
Details
This query command is used to read the result of the CALC 2 operation. If CALC 2 is disabled or
NONE is selected, the "raw" reading is read instead.
Note that this command does not initiate a calculation. This command is used only to read the result
of the last calculation.
Also see
None

CALCulate3

This Calculate subsystem is used to configure and control the limit tests (LIMIT 1 test and LIMIT 2 test).
14-38
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading