Model 6517B Electrometer Reference Manual

A/D controls

:ARSPeed <name>
This command controls autoranging speed.
Usage
:SYSTem:ARSPeed <name>
:ARSPeed?
<name>
Details
This command is used to control autoranging speed. If the autoranging causes the instrument to
rapidly alternate between ranges, changing the speed to NORMal delays measurements after
changing ranges to help avoid oscillating between ranges.
Also see
None
:TSControl <b>
This command controls external temperature readings.
Usage
:SYSTem:TSControl <b>
:TSControl?
<b>
Details
This command controls external temperature readings. When enabled (and the Model 6517-TP is
connected to the instrument), the Model 6517B makes external temperature readings. Note that
ETEMperature has to be a selected data element in order to include the temperature reading in the
data string.
Also see
ETEMperature
6517B-901-01 Rev. C / August 2015
FAST: Fast autoranging
NORMal: Normal autoranging
0 or OFF: Disable external temperature readings
1 or ON: Enable external temperature readings
Section 14: SCPI command reference
14-131