Buffer Multiple Displays - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual
Continuous sequence
Action
STORE
ENTER
RECALL
EXIT
RECALL
EXIT
EXIT
Pretrigger sequence
Action
STORE
ENTER
TRIG
RECALL
EXIT
A manual trigger is used as an example.
Other pretrigger events include GPIB and trigger link.

Buffer multiple displays

Math operations performed on buffered readings are available when readings are recalled. Press
NEXT DISPLAY to view the math operation on the bottom line of front panel display, in the following
order. Note that the appropriate data store element has to be selected in order to get the reading.
RH and ET: relative humidity and external temperature, for example:
Rdg#+00000 RH=35 % ET= 23°C
Note: In order to get these readings, the appropriate sensor has to be connected to the Model
6517B, and it has to be enabled
V-source: V-source output, for example:
Vsource = +0001.000 V
6517B-901-01 Rev. C / August 2015
Result
STORE 00100 READINGS
Storing reading #xx of 100
100 rdgs stored; continuous ON
Rdg#+00000 @Time=+003.903546 sec
...
100 rdgs stored; continuous ON
Rdg#+00000 @Time=+067.709331 sec
...
100 rdgs stored; continuous ON
STORAGE INTERRUPTED
Acquired 100 of 100 readings
(normal reading display)
Result
STORE 00100 READINGS
Waiting for pretrigger event
Storing reading #xx of 50
Storage complete; press RECALL
Rdg#-00050 @Time=-004.999990 sec
...
Rdg#+00000 @Time=+000.000000 sec
...
Rdg#+00049 @Time=+004.899996 sec
(normal reading display)
Annunciator
(* on)
(* off)
Annunciator
(* on)
(* off)
Section 8: Buffer (data store)
8-7