Sense Command Summary - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual

:SENSe command summary

Command
[:SENSe[1]]
:FUNCtion
:FUNCtion?
:DATA
[:LATest]?
:FRESh?
:VOLTage[:DC]
:APERture
:AUTO
:AUTO
:AUTO?
:APERture?
:NPLCycles
:AUTO
:AUTO
:AUTO?
:NPLCycles?
:RANGe
[:UPPer]
[:UPPer]?
:AUTO
:AUTO
:AUTO?
:REFerence
:STATe
:STATe?
:ACQuire
:REFerence?
:DIGits
:AUTO <b>
:AUTO ONCE
6517B-901-01 Rev. C / August 2015
Description
<name>
Select measurement function:
'VOLTage[:DC]', 'CURRent[:DC]',
'RESistance', 'CHARge'
Query function
Path to return readings:
Return the last instrument reading
Query fresh readings
Path to configure volts:
<n>
Set integration rate in seconds (166.67e-6
to 200e-3)
<b>
Enable or disable auto aperture
ONCE
Enable and then disable auto aperture
Query auto aperture
Query aperture (integration rate)
<n>
Set integration rate (line cycles; 0.01 to
10)
<b>
Enable or disable auto NPLC
ONCE
Enable and then disable auto NPLC
Query auto line cycle integration
Query line cycle integration rate
Path to configure measurement range:
<n>
Select range (0 to 210)
Query range
<b>
Enable or disable auto range
ONCE
Set range based on present input signal
:ULIMit
<n>
Set upper limit for auto range (0 to 210)
:ULIMit?
Query upper limit
:LLIMit
<n>
Set lower limit for auto range (0 to 210)
:LLIMit?
Query lower limit
Query auto range
<n>
Specify reference (-210 to +210)
<b>
Enable or disable reference
Query state of reference (0 or 1)
Use input signal as reference
Query reference value
<n>
Specify measurement resolution (4 to 7)
Enable or disable auto resolution
Enable and then disable auto resolution
Section 14: SCPI command reference
Default parameter
SCPI
'VOLT:DC'
(Note 1)
OFF
1
OFF
200
(Note 4)
200
2
0
OFF
6
14-13

Advertisement

Table of Contents
loading