Section 14: SCPI command reference
:TIMer <n>
This command is used to set the interval for the scan layer timer and measure layer.
Usage
:ARM:[SEQuence[1]]:LAYer2:TIMer <n>
:TRIGger:[SEQuence[1]]:TIMer <n>
:TIMer?
:TIMer? DEFault
:TIMer? MINimum
:TIMer? MAXimum
<n>
Details
These commands are used to set the interval for the scan layer and measure layer timers. Note that
the timer is in effect only if the timer is the selected control source. Also, note that the arm layer (Arm
Layer 1) does not use a timer.*
Also see
None

:SIGNal

This command is used to bypass the arm, scan, and measure control sources.
Usage
:ARM[:SEQuence[1]][:LAYer[1]]:SIGNal
:ARM[:SEQuence[1]]:LAYer2:SIGNal
:TRIGger[:SEquence[1]]:SIGNal
Details
These action commands are used to bypass the specified control source and also when you do not
wish to wait for the programmed event. Keep in mind that the instrument must be waiting for the
appropriate event when the command is sent. Otherwise, an error occurs and this command is
ignored.
Also see
None
14-150
0.001 to 999999.999: Specify timer interval in seconds
DEFault: 0.1 second
MINimum: 0.001 second
MAXimum: 999999.999 seconds