Voltage Burden - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 4: Basic measurements
Model 6517B Electrometer Reference Manual

Voltage burden

The input resistance of the ammeter causes a small voltage drop across the input terminals. This
voltage is known as the voltage burden. If the voltage burden is large in relation to the voltage of the
measured circuit, then significant measurement errors occur.
Refer to the figure below to see how voltage burden affects current measurements. Assume VS is 5
mV and RS is 5 k to configure a 1 µA current source (5 mV/5 k = 1 µA). An ideal ammeter with
zero voltage burden would measure the current source as follows:
In practice however, every ammeter has a voltage burden. If the voltage burden (VB) is 1 mV, the
current is measured as follows:
The 1 mV voltage burden caused a 20 % measurement error. Percent error in a measured reading
(IM) due to voltage burden can be calculated as follows:
The voltage burden of the Model 6517B depends on the selected range (see
Accuracy Calculations
(on page A-1)). Voltage burden may be reduced by performing the offset adjustment procedure.
Figure 37: Voltage burden
Noise
Noise can seriously affect sensitive current measurements. The following paragraphs discuss how
source resistance and input capacitance affect noise performance.
Source resistance
The source resistance of the DUT affects the noise performance of current measurements. As the
source resistance is reduced, the noise gain of the ammeter increases.
The following figure shows a simplified model of the feedback ammeter. R
and C
represent the
S
S
source resistance and source capacitance, V
is the source voltage, and V
is the noise voltage.
S
NOISE
Finally, R
and C
are the feedback resistance and capacitance respectively.
F
F
The source noise gain of the circuit can be given by the following equation:
4-14
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading

Table of Contents