Voltage Measurement Ranges Calibration; Voltage Source Calibration - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual
Model 6517B
coulombs
range
2 nC

Voltage measurement ranges calibration

To calibrate the voltage measurement ranges:
1. Connect the DC voltage calibrator to the Model 6517B input.
2. Apply 0.0 V
3. Send the following commands to:
:CAL:PROT:VZERO2
:CAL:PROT:VZERO20
:CAL:PROT:VZERO200
4. Leaving the DC voltage calibrator connected, calibrate each polarity of each range:
b.
Send the calibration command listed in the second column of the below table with the voltage reading
from the DC voltage calibrator.
Model 6517 voltage range calibration commands
Model 6517B
Range/Full-scale
voltage
+2 V
-2 V
+20 V
-20 V
+200 V
-200 V

Voltage source calibration

For all voltage source ranges, connect the DMM to the voltage source output. Set the DMM to
measure DC voltage, and then turn auto ranging on.
6517B-901-01 Rev. C / August 2015
Calibrator
Coulomb zero calibration
voltage
command
2.0 V
:CAL:PROT:CZEROA2N
a.
Apply the voltage listed in the first column of the below table.
Voltage range calibration command
:CAL:PROT:V2 <actual voltage value>
:CAL:PROT:VN2 <actual voltage value>
:CAL:PROT:V20 <actual voltage value>
:CAL:PROT:VN20 <actual voltage value>
:CAL:PROT:V200 <actual voltage value>
:CAL:PROT:VN200 <actual voltage value>
Section 15: Calibration procedure
Coulomb full scale calibration
command
:CAL:PROT:C2N <value>
15-11