Model 6517B Electrometer Reference Manual

Charge calibration commands

When calibrating the 20 mA to 20 mA ranges, use a DC current calibrator. For all current ranges, the
actual current supplied must be within ± 5 % of the nominal value otherwise a calibration error occurs.
Charge calibration commands
Command
:CAL:PROT:CZEROA2N
:CAL:PROT:C2N <+2nC value>
:CAL:PROT:CZEROB2N
:CAL:PROT:CN2N <-2nC value>
:CAL:PROT:CZEROA20N
:CAL:PROT:C20N <+20nC value>
:CAL:PROT:CZEROB20N
:CAL:PROT:CN20N <-20nC value>
:CAL:PROT:CZEROA200N
:CAL:PROT:C200N <+200nC value>
:CAL:PROT:CZEROB200N
:CAL:PROT:CN200N <-200nC value>
:CAL:PROT:CZEROA2U
:CAL:PROT:C2U <+2uC value>
:CAL:PROT:CZEROB2U
:CAL:PROT:CN2U <-2uC value>
6517B-901-01 Rev. C / August 2015
Section 15: Calibration procedure
Description
Start +2 nC range calibration
Calibrate the +2 nC range
Start -2 nC range calibration
Calibrate the -2 nC range
Start +20 nC range calibration
Calibrate the +20 nC range
Start -20 nC range calibration
Calibrate the -20 nC range
Start +200 nC range calibration
Calibrate the +200 nC range
Start -200 nC range calibration
Calibrate the -200 nC range
Start +2 uC range calibration
Calibrate the +2 uC range
Start -2 uC range calibration
Calibrate the -2 uC range
15-19