Input Configurations; Maximum Input Levels - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 3: Connections

Input configurations

The input connector can be configured with guard off or guard on.
With GUARD off, input low is connected to the inner shell of the connector, as shown in the following
figure. This configuration is used for current, resistance, coulombs and unguarded voltage
measurements.
Where possible, make input low connections directly to the INPUT connector low terminal instead of
using COMMON to avoid internal voltage drops that may affect measurement accuracy.
With GUARD on, guard is connected to the inner shell of the triaxial connector, as shown in the
following figure. Input low is accessed through the COMMON connector on the rear panel. This
configuration is used for guarded voltage measurements only. Note that guard can only be enabled
(on) for the volts function. For ohms, amps and coulombs, guard is always disabled (off). For voltage
measurements, guard is enabled or disabled from the Configure Voltage menu structure.
The INPUT triaxial connector is also used for the force voltage measure current configuration. This
configuration uses the V-source to make resistance measurements and current measurements.

Maximum input levels

The maximum input levels to the Keithley Instruments Model 6517B Electrometer are summarized in
the figure below.
The maximum common-mode input voltage (the voltage between input low and chassis
ground) is 500 V
3-2
Figure 9: Input connector configuration - unguarded
Figure 10: Input connector configuration - guarded
. Exceeding this value may create a shock hazard.
peak
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015