Model 6517B Electrometer Reference Manual
:NPLCycles <n>
This command is used to set NPLC for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:NPLCycles
[:SENSe[1]]:CURRent[:DC]:NPLCycles
[:SENSe[1]:RESistance:NPLCycles
[:SENSe[1]]:CHARge:NPLCycles
:NPLCycles?
:NPLCycles?
:NPLCycles?
:NPLCycles?
<n>
Details
The integration period (measurement speed) for the Model 6517B can be set using either of two
commands; :APERture or :NPLCycle. Aperture specifies time (in seconds) per integration, while
NPLC expresses the integration period by basing it on the power line frequency.
When the integration period is set using the :NPLCycles command, the time value for the
:APERture command changes accordingly to reflect the new integration rate. Conversely, if the
integration period is set using the :APERture command, the value for the :NPLCycle command
changes accordingly.
On power-up, the instrument uses the NPLC value to determine the integration period.
Also see
:APERture <n>
6517B-901-01 Rev. C / August 2015
<n>
0.01 to 10: Power line cycles per integration
DEFault: 1
MINimum: 0.01
MAXimum: 10
(on page 14-59)
<n>
<n>
<n>