Model 6517B Electrometer Reference Manual
:AUTO <b>|ONCE
This command is used to control the auto resolution for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:DIGits:AUTO <b>|ONCE
[:SENSe[1]]:CURRent[:DC]:DIGits:AUTO <b>|ONCE
[:SENSe[1]:RESistance:DIGits:AUTO <b>|ONCE
[:SENSe[1]]:CHARge:DIGits:AUTO <b>|ONCE
:AUTO?
<b>
Details
Description These commands are used to enable or disable auto resolution for the specified function.
With auto resolution selected, the instrument automatically selects the optimum resolution for the
present integration time setting.
The ONCE parameter is analogous to a momentary toggle switch. When :AUTO ONCE is sent, auto
resolution turns on momentarily and then disables. While enabled, resolution is selected
automatically. Auto resolution then disables but the selected value is retained.
The auto resolution command (:DIGits:AUTO) is coupled to the command that manually selects
resolution (:DIGits <n>). When auto resolution is enabled, the parameter value for :DIGits <n>
changes to the automatically selected resolution value. Thus, when auto resolution is disabled, the
instrument remains at the automatically selected resolution.
Also see
:DIGits <n>

:AVERage commands

The :AVERage commands are used to configure and control the digital filter
information).
6517B-901-01 Rev. C / August 2015
1 or ON: Enable auto resolution
0 or OFF: Disable auto resolution
ONCE: Enable and then disable auto resolution
(on page 14-74)
Section 14: SCPI command reference
(Filters
(on page 9-1) for more
14-75