Model 6517B Electrometer Reference Manual

Error messages

The following error messages are associated with RS-232 interface operation. These error messages
are also listed in
Bit B11 of the Operation Condition Register.
Error messages
Error code
+800
+801
+802
+803
+804
+805
+806
+807
+808
6517B-901-01 Rev. C / August 2015
Status and error messages
Error message
RS-232 framing error detected
RS-232 parity error detected
RS-232 overrun detected
RS-232 break detected
RS-232 noise detected
Invalid system communication
RS-232 settings lost
RS-232 OFLO; characters lost
ASCII only with RS-232
(on page 2-11). Note that any of the following errors sets
Description
Start or stop bit mismatch between computer and
Model 6517B.
Parity type mismatch between computer and Model
6517B.
Data received before previous data can be read.
A break occurred (device-clear occurs).
Noisy signal that could corrupt data.
RS-232 interface selection lost on power-up;
instrument defaults to GPIB.
RS-232 settings lost on power-up.
A character was received but discarded due to a lack
of input buffer space. Each line-feed character
(<LF>) closes a buffer and opens another. Thus, if
too many line-feeds are sent in succession, then all
buffers are filled before they have a chance to be
read.
A data format other than ASCII was selected.