Section 14: SCPI command reference
Command
:AUTO?
:DIGits?
:AVERage
:TYPE <name>
:TYPE?
:TCONtrol <name> Select filter type: (MOVing or REPeat)
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe]
[:STATe]?
:MEDian
[:STATe] <b>
[:STATe]?
:RANK <NRf>
:RANK?
:GUARd <b>
:GUARd?
:XFEedback <b>
:XFEedback?
:CURRent[:DC]
:APERture
:AUTO
:AUTO
:AUTO?
:APERture?
:NPLCycles
:AUTO
:AUTO
:AUTO?
:NPLCycles?
:RANGe
[:UPPer]
[:UPPer]?
:AUTO
:AUTO
14-14
Description
Query auto resolution
Query resolution
Path to control the average filter:
Select filter type (NONE, SCALar,
ADVanced)
Query filter type
Query filter type
Specify filter count (1 to 100)
Query filter count
Path to configure and control advanced
filter
Specify noise tolerance level (0 to 100 %)
Query noise tolerance level
<b>
Enable or disable digital filter
Query state of digital filter
Path to control the median filter:
Enable or disable median filter
Query state of median filter
Specify rank for median filter (1 to 5)
Query rank
Enable or disable guard
Query state of guard
Enable or disable external feedback
Query state of external feedback
Path to configure Amps:
<n>
Set integration rate in seconds (166.67e-6
to 200e-3)
<b>
Enable or disable auto aperture
ONCE
Enable and then disable auto aperture
Query auto aperture
Query aperture (integration rate)
<n>
Set integration rate (line cycles; 0.01 to
10)
<b>
Enable or disable auto NPLC
ONCE
Enable and then disable auto NPLC
Query auto line cycle integration
Query line cycle integration rate
Path to configure measurement range:
<n>
Select range (0 to 21e-3)
Query range
<b>
Enable or disable auto range
Set range based on present input signal
ONCE
Model 6517B Electrometer Reference Manual
Default parameter
SCALar
(Note 3)
10
1
OFF
ON
1
OFF
OFF
(Note 1)
OFF
1
OFF
20e-3
(Note 4)
6517B-901-01 Rev. C / August 2015