Model 6517B Electrometer Reference Manual
:STIMe <NRf>
This command specifies the bias time for the specified test.
Usage
:TSEQuence:SRESistivity:STIMe <NRf>
:TSEQuence:VRESistivity:STIMe <NRf>t
:TSEQuence:SIResistance:STIMe <NRf>
:TSEQuence:STSWeep:STIMe <NRf>
:STIMe?
<NRf>
Details
This command is used to specify the bias time for the specified test.
This configuration command is used for the following tests:
Surface resistivity test
Volume resistivity test
Surface insulation resistance test
Staircase sweep test
:TSEQuence:SRESistivity:STIMe specifies the bias time for the surface resistivity test.
:TSEQuence:VRESistivity:STIMe specifies the bias time for the volume resistivity test.
:TSEQuence:SIResistance:STIMe specifies the bias time for the surface insulation resistance
test.
:TSEQuence:STSWeep:STIMe specifies the bias time for the staircase sweep test
:STIMe? queries the bias time.
Also see
None
6517B-901-01 Rev. C / August 2015
0 to 99999.9 delay in seconds (staircase sweep test)
0 to 99999.9 Delay in seconds (all other tests)
Section 14: SCPI command reference