Output1 Subsystem; [:State] <B; Output2 Subsystem; Lsense <Name - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference

:OUTPut1 subsystem

[:STATe] <b>
This command is used to enable or disable the V-source.
Usage
:OUTPut1:[:STATe] <b>
[:STATe]?
<b>
Details
This command is used to control the state of the V-source.
When enabled, the V-source is placed in operate. When disabled, the V-source is placed in standby.
The V-source is configured from the :SOURce subsystem.
Note that the V-source for Auto-V-source Ohms is controlled from the :SENSe Subsystem.
Also see
None

:OUTPut2 subsystem

:LSENse <name>
This command is used to set the polarity of lines #1 through #4.
Usage
:OUTPut2:TTL[1]:LSENse <name>
:OUTPut2:TTL2:LSENse <name>
:OUTPut2:TTL3:LSENse <name>
:OUTPut2:TTL4:LSENse <name>
:LSENse?
<name>
Details
This command is used to set the polarity of the digital output lines. When set for active high (AHIGh)
polarity, the specified output line is true (ON) when the output level is high. The output line is false
(OFF) when the output level is low. When set for active low (ALOW) polarity, the output line is true (ON)
when the output level is low. The output line is false (OFF) when the output level is high.
The logic levels (true or false) of the digital output lines are set from the SOURce Subsystem.
Also see
None
14-56
O or OFF: Disable V-source (standby)
1 or ON: Enable V-source (operate)
AHIGh: Sets polarity of specified line to active high
ALOW: Sets polarity of specified line to active low
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading

Table of Contents