Stop <Nrf; Step <Nrf - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference
:STOP <NRf>
These commands specify the stop test voltages and query the stop voltage for the specific test.
Usage
:TSEQuence:DLEakage:STOP <NRf>
:TSEQuence:STSWeep:STOP <NRf>
:STOP?
<NRf>
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
This command is used to specify the stop voltage for the specified test.
:TSEQuence:DLEakage:STOP <NRf> stops voltage; diode leakage test.
:TSEQuence:STSWeep:STOP <NRf> stops voltage; staircase sweep test.
:STOP? queries the stop voltage.
Also see
None
:STEP <NRf>
This command specifies the step voltage for a diode leakage or staircase sweep test.
Usage
:TSEQuence:DLEakage:STEP <NRf>
:TSEQuence:STSWeep:STEP <NRf>
:STEP?
<NRf>
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
:TSEQuence:DLEakage:STEP specifies the step voltage for a diode leakage test.
:TSEQuence:STSWeep:STEP specifies the step voltage for a staircase sweep test.
:STEP? queries the step voltage
Also see
None
14-160
-100.0 to 100.0 Volts; 100 V range
-1000 to 1000 Volts; 1000 V range
-100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading

Table of Contents