Llevel <Nrf; Ltime <Nrf - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference
:LLEVel <NRf>
This command specifies the low-level voltage for the square wave sweep test.
Usage
:TSEQuence:SQSWeep:LLEVel <NRf>
:LLEVel?
<NRf>
Details
This configuration command is used for the square wave sweep test.
:LLEVel? queries low-level voltage.
Also see
None
:LTIMe <NRf>
This command specifies the low-level time for the square wave sweep test.
Usage
:TSEQuence:SQSWeep:LTIMe <NRf>
:LTIMe?
<NRf>
Details
This configuration command is used for the square wave sweep test.
:LTIMe? queries low-level time.
Also see
None
14-168
-100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
0 s to 9999.9 s
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading

Table of Contents