Model 6517B Electrometer Reference Manual
:LLIMit <n>
This command is used to set the lower limit for volts and amps.
Usage
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO:LLIMit <n>
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO:LLIMit <n>
:ULIMit?
DEFault
MINimum
MAXimum
:LLIMit?
DEFault:LLIMit?
MINimum:LLIMit?
MAXimum
<n>
Details
Automatic range changes slow down the reading rate. By setting upper and lower measurement
range limits, you can prevent inadvertent range changes to ranges that you are not going to use.
The upper range limit is selected by specifying the maximum expected reading that you expect to
measure. For example, if the maximum expected reading is approximately 1mA, let the parameter
(<n>) = 0.001 (or 1e-3). The lowest (most sensitive) range that can measure 1 mA is the 2 mA range.
Thus, the 2 mA range is selected as the upper range limit. With this upper limit, the instrument cannot
uprange to the 20 mA measurement range.
The lower range limit is selected by specifying the lowest expected reading that you expect to
measure.
Also see
:LGRoup <name>
6517B-901-01 Rev. C / August 2015
Query largest allowable lower limit
0 to 210: Range limit for volts
0 to 21e-3: Range limit for amps
DEFault : 210 (volts)
MINimum: 0 (volts and amps)
MAXimum: Same as DEFault
(on page 14-66)
21e-3 (amps)