Vsource:range <N - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual
:VSOurce:RANGe <n>
This command is used to set the V-source range for Manual V-source Ohms.
Usage
[:SENSe[1]:RESistance:MANual:VSOurce:RANGe <n>
:RANGe?
:RANGe?
:RANGe?
:RANGe?
<n>
Details
This command is used to select the optimum V-source range for Auto V-source Ohms. Specifying the
output voltage value automatically selects the most sensitive range. For example, specifying a value
less than or equal to 100 selects the 100 V range. Specifying a value greater than 100 selects the
1000 V range.
You cannot select the 100 V range if the programmed amplitude is greater than 100 V; a Settings
Conflict error results.
Also see
:VSOurce[:AMPLitude] <n>
6517B-901-01 Rev. C / August 2015
DEFault
MINimum
MAXimum
0 to 1000: Specify output in volts
DEFault: 100V range
MINimum: 100V range
MAXimum: 1000V range
(on page 14-70)
Section 14: SCPI command reference
14-69

Advertisement