Poflag <Name; Abort - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference
:POFLag <name>
This command is used to control the no-operation-pending flag.
Usage
:INITiate:POFLag <name>
:POFLag?
<name>
Details
This command is used to include or exclude the no-operation-pending flag in the internal process of
determining whether all pending operations of an initiate command are completed.
Also see
*OPC (operation complete)
*OPC? (operation complete query)

:ABORt

This command is used to abort the operation.
Usage
:ABORt
Details
When this action command is sent, the Model 6517B aborts the operation and returns to the top of
the trigger model. If continuous initiation is disabled, the instrument goes the idle state. If continuous
initiation is enabled, operation continues on into Arm Layer 1.
The abort command resets the source bypasses of the trigger model. If scanning, the scan pointer
resets back to the first channel in the scan list.
Also see
None
14-146
INCLude: Include no-operation-pending flag
EXCLude: Exclude no-operation-pending flag
(on page 12-8)
(on page 12-9)
Model 6517B Electrometer Reference Manual
6517B-901-01 Rev. C / August 2015

Advertisement

Table of Contents
loading

Table of Contents