Temperature Calibration Commands - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual

Temperature calibration commands

This is a list of all Model 6517B temperature calibration commands.
Temperature calibration commands
Command
:CAL:PROT:TZERO
:CAL:PROT:T100 0.004096
:CAL:PROT:TN25 -0.000968
Miscellaneous calibration commands
This is a list of the remaining Model 6517B calibration commands.
Miscellaneous calibration commands
Command
:CAL:PROT:COUNT?
:CAL:PROT:DATE yyyy,mm,dd
:CAL:PROT:DATE?
:CAL:PROT:NDUE yyyy,mm,dd
:CAL:PROT:NDUE?
:CAL:PROT:CODE 'KInnnnnn'
:CAL:PROT:SAVE
:CAL:PROT:LOCK
6517B-901-01 Rev. C / August 2015
Description
Calibrate the temperature input zero
Calibrate the temperature 100 °C range
Calibrate the temperature -25 °C range
Description
Query the number of times the instrument has been
calibrated.
Set the calibration date
Query the calibration date
Set the calibration due date
Query the next calibration due date
Unlock the calibration using the password (default is
006517)
Save the calibration
Lock the calibration
Section 15: Calibration procedure
15-23

Advertisement

Table of Contents
loading

Table of Contents