Model 6517B Electrometer Reference Manual
:TCONfigure commands
:PROTocol <name>
This command is used to specify the trigger link protocol.
Usage
:TRIGger[:SEQuence[1]]:TCONfigure:PROTocol <name>
:PROTocol?
<name>
Details
This command is used to select the protocol for the measure layer trigger link. With ASYNchronous
selected, separate trigger lines are used for input and output triggers. With SSYNchronous selected,
a single trigger line is used for both input and output triggers.
Note that the arm and scan layers only use the asynchronous trigger link mode and therefore, do not
have to be selected.
Also see
None
6517B-901-01 Rev. C / August 2015
ASYNchronous: Asynchronous trigger link mode
Synchronous: Semi-synchronous trigger link mode
Section 14: SCPI command reference