[:State] <B - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual
[:STATe] <b>
This command
Usage
[:SENSe[1]]:VOLTage[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:CURRent[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:RESistance:MEDian[:STATe] <b>
[:SENSe[1]]:CHARge:MEDian[:STATe] <b>
[STATe]?
<b>
Details
This command is used to control (on or off) the median filter. When enabled, the instrument
determines and displays the median reading for the specified number of readings. The :RANK
command is used to specify the number of readings used to determine the median.
Note that the median filter can be used with the average filter. When both of these filters are enabled,
each displayed reading reflects the median for the average readings.
Setting STATe to OFF with AVER:TYPE set to NONE sets AVER:STAT to OFF.
Also see
None
6517B-901-01 Rev. C / August 2015
1 or ON: Enable median filter
0 or OFF: Disable median filter
Section 14: SCPI command reference
14-79