Mtime <Nrf - Keithley 6517B Reference Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 14: SCPI command reference
:MTIMe <NRf>
This command specifies the measure time for the specified test.
Usage
:TSEQuence:SRESistivity:MTIMe <NRf>
:TSEQuence:VRESistivity:MTIMe <NRf>
:TSEQuence:SIResistance:MTIMe <NRf>
:TSEQuence:ALTPolarity:MTIMe <NRf>
:MTIMe?
<NRf>
Details
This command is used to specify the measure time for the specified test.
This configuration command is used for the following tests:
Surface resistivity test
Volume resistivity test
Surface insulation resistance test
Alternating polarity resistance/resistivity test
:TSEQuence:SRESistivity:MTIMe specifies the measure time for the surface resistivity test.
:TSEQuence:VRESistivity:MTIMe specifies the measure time for the volume resistivity test.
:TSEQuence:SIResistance:MTIMe specifies the measure time for the surface insulation
resistance test.
:TSEQuence:ALTPolarity:MTIMe specifies the measure time for the alternating polarity test.
:MTIMe? queries the measure time of the specified test.
Also see
None
14-166
0 to 9999.9 (seconds)
0.5 to 9999.9 (seconds) for the alternating polarity resistance/resistivity test
Model 6517B Electrometer Reference Manual