Model 6517B Electrometer Reference Manual
:VSControl <name>
This command is used to select the ohms V-source mode.
Usage
[:SENSe[1]]:RESistance:VSControl <name>
:VSControl?
<name>
Details
This command is used to select Manual V-source ohms or Auto V-source ohms. With MANual
selected, you can set the V-source level for ohms measurements.
With AUTO selected, the V-source level is automatically set for the ohms measurement (40V or 400V).
Also see
None
:MSELect <name>
This command is used to select the ohms measurement type.
Usage
[:SENSe[1]]:RESistance:MSELect <name>
:MSELect?
<name>
Details
This command is used to select the ohms measurement type. NORMal is used to make normal resistance
measurements, and RESistivity is used for resistivity (surface and volume) measurements (see
measurements
(on page 4-22) for more information).
Also see
None

:RESistivity commands

The following commands are used for resistivity measurements. If using the Model 8009 Resistivity
Test Fixture, the measurement type (surface or volume) is automatically sensed by the Model 6517B
through the safety interlock cable. Also, when using the Model 8009, you do not need to use the
:USER commands (:RSELect, :KSURface, and :KVOLume).
See also
6517B-901-01 Rev. C / August 2015
MANual: Manual V-source Ohms
AUTO: V-source Ohms
NORMal: Normal resistance measurements
RESistivity: Resistivity measurements
Resistivity measurements
(on page 4-22) for more information.
Section 14: SCPI command reference
Resistivity
14-85