Keithley 6517B Reference Manual page 270

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Model 6517B Electrometer Reference Manual
Command
:TYPE <name>
:TYPE?
:TCONtrol
<name>
:TCONtrol?
:COUNt
:COUNt?
:ADVanced
<n>
[:STATe]
[:STATe]?
:MEDian
[:STATe] <b>
[:STATe]?
:RANK <NRf>
:RANK?
:ADIScharge
[:STATe] <b>
[:STATe]?
:LEVel <NRf>
:LEVel?
NOTES:
1. 16.67 ms for 60 Hz power line frequency and 20ms for 50 Hz line frequency.
2. *RST — OFF, :SYSTem:PRESet — ON.
3. *RST — REPeat, :SYSTem:PRESet — MOVing.
4. *RST — ON, :SYSTem:PRESet — OFF.
6517B-901-01 Rev. C / August 2015
Description
Select filter type (NONE, SCALar,
ADVanced)
Query filter type
Select filter type: (MOVing or REPeat)
Query filter type
<n>
Specify filter count (1 to 100)
Query filter count
Path to configure and control advanced
filter
:NTOLerance
Specify noise tolerance level (0 to 100 %)
:NTOLerance?
Query noise tolerance level
<b>
Enable or disable filter
Query state of digital filter
Path to control the median filter:
Enable or disable median filter
Query state of median filter
Specify rank for median filter
Query rank
Path to control auto discharge:
Enable or disable auto discharge
Query state of auto discharge
Specify level (-2.2e6 to 2.2e6)
Query level
Section 14: SCPI command reference
Default parameter
SCPI
NONE
MOVing
10
1
OFF
ON
1
OFF
2e-6
14-19